Start • Knowledge Pathway • Tutorials • When to use a Vibrating Microtome for Tissue Sectioning When to use a Vibrating Microtome for Tissue Sectioning Users of microtomes often ask if they can use just one microtome across all functions in the laboratory. This article addresses the advantages and limitations of Vibrating Blade microtomes. Other microtome types require hardening the tissue by freezing or paraffin embedding. This is because a blade, no matter how sharp, pushed forward into a block of soft tissue, will compress the tissue in front of it rather than cut. When compressed enough to push back, the tissue will suddenly cut, but not straight through, the blade bevel and uneven compression will create deflections. However, soft tissue can be cut by rapidly moving the blade back and forth, thus creating a sawing effect as the blade slowly advances forward. Why cut soft? Advantages of a Vibrating Microtome: Does not fracture cell membranes (freezing frequently does); cytosol remains in cells. Stains are sharper, with no cytosol in the extracellular tissue. Better for HRP, etc. Can keep cells living. Tissue may be oxygenated under media while cutting. Quick setup, glue tissue to the platform. No encapsulation is needed. Disadvantages of a Vibrating Microtome: Much slower to operate than a cryostat or rotating or sliding microtome. Sections must be thick, 50 microns and up. Block from which sections are taken must not be too thick. Issues encountered in Vibrating Microtomy. Blade Angle: The blade can be set to any angle. The correct blade angle depends only on the properties of the blade. However, the consequences of incorrect blade angle are different depending on the properties of the material being cut as illustrated below. Image This image from an old manual actually shows an incorrect blade angle and clearance angle. See the article Knife Angle in Microtomy. The "Blade Angle" is the angle between the blade body and the direction of motion. Note that there is a bevel on the front edge of the blade. This is necessary for blade manufacturing. The clearance angle is the angle between the lower bevel and the direction of motion. The blade angle forces the tissue to bend upward at the blade front. This compresses cells on one side of the section at the bend and stretches them on the other. This distortion should always be minimized. However, if the blade angle is zero, the clearance angle will be negative because the bevel is still there. The back of the lower bevel will push down and compress the block as it passes over and slide along the tissue. This is highly undesirable for any application. To avoid this risk, the clearance angle, the angle between the bevel on the blade and the path of advance, must be above zero. At zero, there would still be some friction. An angle of about 1° would be perfect. There is no case where more than that is advantageous. Note in the imagethat it is about 10-15 degrees, and thus bending the tissue more than is necessary. The correct blade angle is thus dependent on, and slightly greater than, the bevel angle of the blade being used, and no other property. Amplitude of Oscillation For mechanical reasons having to do with the natural frequency of oscillation, it is not mechanically advisable to speed up the back and forth motion over the same range. However, by increasing the amplitude of oscillation, with each oscillation still at the same frequency, one can have the blade moving faster against the tissue part of the time. As long as the greater amplitude is not tearing the tissue at the extremes or causing vertical vibration of the blade, more is better, especially for tissue that wants to compress. Speed of Forward Advance If the blade is slipping out of the tissue block or compressing the tissue, try a slower advance. For the most critical applications, such as calcium dual fluorescent imaging in live tissue sections, the slower, the better, some use 10 minutes per section, as long as the tissue is still alive when you have the section. The trade-off is the speed of getting the job done .versus section quality. Thick or Thin Sections If the instrument is cutting alternate sections of different thickness, there are a few possible causes. Clearance angle negative is a common one; the blade angle is too low. Another is something loose; the pedestal, the tissue on it, or the blade is not tightly secured. Try to grasp and wobble things one at time, preferably without cutting yourself. Another possible cause is that the tissue has a high height for the base of support, and is flexing at the top, so the tissue is moving. Cut the block smaller if feasible or find a way to support the tissue better. Die Inhalte des Knowledge Pathway von Leica Biosystems unterliegen den Nutzungsbedingungen der Website von Leica Biosystems, die hier eingesehen werden können: Rechtlicher Hinweis. Der Inhalt, einschließlich der Webinare, Schulungspräsentationen und ähnlicher Materialien, soll allgemeine Informationen zu bestimmten Themen liefern, die für medizinische Fachkräfte von Interesse sind. Er soll explizit nicht der medizinischen, behördlichen oder rechtlichen Beratung dienen und kann diese auch nicht ersetzen. Die Ansichten und Meinungen, die in Inhalten Dritter zum Ausdruck gebracht werden, spiegeln die persönlichen Auffassungen der Sprecher/Autoren wider und decken sich nicht notwendigerweise mit denen von Leica Biosystems, seinen Mitarbeitern oder Vertretern. Jegliche in den Inhalten enthaltene Links, die auf Quellen oder Inhalte Dritter verweisen, werden lediglich aus Gründen Ihrer Annehmlichkeit zur Verfügung gestellt. Vor dem Gebrauch sollten die Produktinformationen, Beilagen und Bedienungsanleitungen der jeweiligen Medikamente und Geräte konsultiert werden. Copyright © 2024 Leica Biosystems division of Leica Microsystems, Inc. and its Leica Biosystems affiliates. All rights reserved. LEICA and the Leica Logo are registered trademarks of Leica Microsystems IR GmbH. Wenn Sie sich dieses Schulungswebinar aus unserer monatlichen Reihe angesehen haben und dessen Anerkennung als Weiterbildung seitens der Sie zertifizierenden Organisation beantragen möchten, füllen Sie bitte das Formular aus, um Ihrem Protokoll die entsprechenden Weiterbildungspunkte gemäß Ihren eigenen Angaben hinzuzufügen. ANMELDUNG FÜR WEITERBILDUNGSPUNKTE NACH EIGENANGABEN Erhalten Sie neueste Informationen aus dem Knowledge Pathway direkt in Ihr Postfach. Melden Sie sich jetzt an! Wir sind stets auf der Suche nach weiteren interessanten Autoren, die wir hier vorstellen möchten. Reichen Sie Ihren Beitrag ein und wir melden uns bei Ihnen! Senden Sie Ihren Beitrag SHARE Facebook Twitter LinkedIn Email